← FDA ENFORCEMENT
Z-2064-2026Class II

Erbe Medical, LLC

Recalling firm
Erbe Medical, LLC
Product type
Device
Status
Ongoing
Report date
2026-05-13
Distribution
US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
Firm state
AZ
Product quantity
5,500
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

Product description

ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222

Code / lot info

UDI-DI: 04065655000252. Lots: W4465580, W4466022, W4466023, WO469442, WO469923, WO473978, WO473979, WO473980, WO473981, WO476347, WO477013, WO477014

Other recalls SPEQ maps to Environmental Monitoring
2026-07-22Class IFrutas y Hortalizas del Sur S.A.2026-07-22Class IIGELLERT GLOBAL GROUP2026-07-22Class IIChiesi USA, Inc.2026-07-22Class IICareFusion 213, LLC
See all Environmental Monitoring recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.