FDA PRODUCT CODE · CLASS II
OIK — Radiology Dental Tray
FDA has exempted these devices from the premarket notification requirement by publishing in the Federal Register of July 11, 2017 (82 FR 31976) available at https://www.federalregister.gov/documents/2017/07/11/2017-14453/medical-devices-exemptions-from-premarket-notification-class-ii-devices subject to the limitations under 21 CFR XXX.9, where XXX refers to Parts 862-892, and any limitations on the exemption described in (82 FR 31976). Any device that exceeds the limitations found in 21 CFR sections 862.9 to 892.9 or identified in the Federal Register (82 FR 31976) available at https://www.federalregister.gov/documents/2017/07/11/2017-14453/medical-devices-exemptions-from-premarket-notification-class-ii-devices is required to submit a premarket notification submission (510(k)) to FDA before introducing a device into interstate commerce. If a manufacturers device is exempt from the premarket notification requirement, then a premarket notification submission and FDA clearance is not required before marketing the device in the U.S. however, manufacturers are required to meet all other requirements under the Federal Food Drug & Cosmetic and implementing regulations including (but not limited to) registration and listing requirements. Please see the Device Registration and Listing website for additional information on registration and listing. This product code has been established in accordance with the may 20, 1997, guidance entitled, convenience kits interim regulatory guidance, found at https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm080217.pdf. This type of convenience kit, as listed in the guidance above, is under enforcement discretion, and does not require a premarket notification (510(k)) to market if it meets all criteria in the guidance.
What these counts are not
Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.
What this classification requires
Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.
510(K) Exempt
SPEQ READINGNo premarket notification is required for a device in this code, within the limitations of exemption. Exempt from premarket review is NOT exempt from regulation: registration, listing, labelling, adverse-event reporting and — unless the code is also GMP exempt — 21 CFR Part 820 all still apply.
Class II (moderate to high risk): general controls and special controls
SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.
Device is eligible for The Voluntary Malfunction Summary Reporting Program
SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.
Not exempt due to Good Manufacturing Practice (GMP)/Quality System
SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.
Device is not implantable
SPEQ READINGNo implant-specific expectations attach from the classification itself.
Device is not used for life sustaining purposes
SPEQ READINGNo life-sustaining designation attaches to this code.
Device is not a candidate for a third party review program
SPEQ READINGFDA reviews any premarket notification for this code directly.
Other product codes under 21 CFR 892.1650
One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.
21 CFR 892.1650 — Image-intensified fluoroscopic x-ray system
The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.
(a) Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
(b) Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. [53 FR 1567, Jan. 20, 1988, as amended at 66 FR 57369, Nov. 15, 2001; 84 FR 71818, Dec. 30, 2019]
Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 892.1650 on eCFR →
Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →