FDA PRODUCT CODE · CLASS II

LZO — Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 888.3353
MEDICAL SPECIALTY
Orthopedic
REVIEW PANEL
OR
✓ Implant— Life-sustaining— GMP-exempt— Third-party review eligible
26
510(K)/PMA CLEARANCES
510k: 26
76
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is implantable

SPEQ READINGImplant status raises the evidentiary bar on biocompatibility and on traceability from finished device back to material lot.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is not a candidate for a third party review program

SPEQ READINGFDA reviews any premarket notification for this code directly.

Other product codes under 21 CFR 888.3353

One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.

21 CFR 888.3353 — Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A1 2 0 3 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

(b) Classification. Class II. [54 FR 48239, Nov. 22, 1989; 54 FR 51342, Dec. 14, 1989]

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 888.3353 on eCFR →

Recent clearances

K262473Microport Orthopedics, Inc. · 510k2026-10-01
K260173Medacta International S.A. · 510k2026-09-17
K253948United Orthopedic Corporation · 510k2026-09-02
K261009Waldemar Link GmbH & Co. KG · 510k2026-07-06
K260831Zimmer, Inc. · 510k2026-06-16
K260182Zimmer, Inc. · 510k2026-04-27
K260037Implantcast GmbH · 510k2026-03-30
K252401Implantcast GmbH · 510k2026-03-12

Recent recalls

BioPro, Inc.Open, Classified · 2025-10-14

Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.

Waldemar Link GmbH & Co. KG (Mfg Site)Open, Classified · 2024-12-26

Error on the main label and the patient label. The labels indicate a size XXL, this is a standard-size neck segment.. All further information on the labels is correct (Length, CCD and Offset).

Biomet, Inc.Open, Classified · 2023-12-18

One product complaint was received reporting that there was an incorrect adapter in the packaging. An adapter with neck length M/+0 was inside the product packaging, which should have contained an adapter with neck length S/-3.0. The difference in neck length may be recognized by the size (S or M) i

Biomet, Inc.Open, Classified · 2023-12-18

One product complaint was received reporting that there was an incorrect adapter in the packaging. An adapter with neck length M/+0 was inside the product packaging, which should have contained an adapter with neck length S/-3.0. The difference in neck length may be recognized by the size (S or M) i

Corin LtdOpen, Classified · 2022-12-19

The BIOLOX Delta Ceramic Head size 36XL is labelled as the size 32XL and vice-versa.

Corin LtdOpen, Classified · 2022-12-19

The BIOLOX Delta Ceramic Head size 36XL is labelled as the size 32XL and vice-versa.

MicroPort Orthopedics Inc.Terminated · 2017-01-03

The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.

MicroPort Orthopedics Inc.Terminated · 2016-12-01

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one perce

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →