LZH — Pump, Infusion, Enteral
What these counts are not
Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.
What this classification requires
Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.
510(K)
SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.
Class II (moderate to high risk): general controls and special controls
SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.
Device is not eligible for The Voluntary Malfunction Summary Reporting Program
SPEQ READINGIndividual malfunction reporting under 21 CFR Part 803 applies. The complaint-handling system has to produce a decision per event, not per period.
Not exempt due to Good Manufacturing Practice (GMP)/Quality System
SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.
Device is not implantable
SPEQ READINGNo implant-specific expectations attach from the classification itself.
Device is not used for life sustaining purposes
SPEQ READINGNo life-sustaining designation attaches to this code.
Device is not a candidate for a third party review program
SPEQ READINGFDA reviews any premarket notification for this code directly.
Other product codes under 21 CFR 880.5725
One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.
21 CFR 880.5725 — Infusion pump
The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.
(a) Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
(b) Classification. Class II (performance standards).
Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 880.5725 on eCFR →
Recent clearances
Recent recalls
The firm is adding an addendum to the operator's manual for Infinity Enteralite Ambulatory Feeding Pumps pertaining to the Infinite Dose mode and the inability of the Pump to detect air in line.
MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 m
Corpak is removing the Corflo 300 Enteral Pump from the market due to the possibility that with fluid ingress or other damage, the keypad may not function, and the possibility of motor mount failure.
Pumps were manufactured with an incorrect circuit board that may cause an increase in the feed rate and flush frequency of fluids.
Some of the sets may leak or disconnect near the pump adapter during use. The patient could ingest air and vomit causing a safety concern.
Some of the sets may leak or disconnect near the pump adapter during use. The patient could ingest air and vomit causing a safety concern.
Some of the sets may leak or disconnect near the pump adapter during use. The patient could ingest air and vomit causing a safety concern.
The feeding spike sets may leak at the interface of the tube and spike connector.
Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →