FDA PRODUCT CODE · CLASS II

LHI — Set, I.V. Fluid Transfer

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 880.5440
MEDICAL SPECIALTY
General Hospital
REVIEW PANEL
HO
— Implant— Life-sustaining— GMP-exempt✓ Third-party review eligible
9
510(K)/PMA CLEARANCES
510k: 9
28
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is not implantable

SPEQ READINGNo implant-specific expectations attach from the classification itself.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is a candidate for a third party review program

SPEQ READINGAn Accredited Person may conduct the 510(k) review instead of FDA. FDA still makes the clearance decision — third-party review changes who reviews, never who decides.

Other product codes under 21 CFR 880.5440

One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.

21 CFR 880.5440 — Intravascular administration set

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

(b) Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9. [45 FR 69682, Oct. 21, 1980, as amended at 66 FR 15798, Mar. 21, 2001]

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 880.5440 on eCFR →

Recent clearances

K252987Epic Medical Pte. , Ltd. · 510k2026-06-12
K260301Hangzhou Qiantang Longyue Biotechnology Co., Ltd. · 510k2026-06-05
K250345Medline Industries, LP · 510k2025-10-24
K251715Asset Medikal · 510k2025-10-14
K251676Microtek Medical, LLC · 510k2025-09-16
K243404Takeda Pharmaceuticals · 510k2025-07-18
K240761Yukon Medical, LLC · 510k2025-03-21
K243985Icu Medical, Inc. · 510k2025-01-22

Recent recalls

West Pharmaceutical Services AZ, Inc.Open, Classified · 2024-11-01

Due to packaging damage from heat exposure during shipping/transportation that may result in a breach in the sterile barrier.

Baxter Healthcare CorporationOpen, Classified · 2024-02-16

An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volume.

Baxter Healthcare CorporationOpen, Classified · 2024-02-16

An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volume.

International Medical Industries, Inc.Open, Classified · 2023-01-07

Blister package had an unsealed edge compromising sterility.

WEST PHARMA. SERVICES IL, LTDTerminated · 2019-01-28

Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag is recalled due to similarities with 13mm Vial2Bag DC.

WEST PHARMA. SERVICES IL, LTDTerminated · 2019-01-28

Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women. There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections

WEST PHARMA. SERVICES IL, LTDTerminated · 2019-01-28

Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag DC 20mm is recalled due to similarities with 13mm Vial2Bag DC.

Baxter Corporation EnglewoodTerminated · 2016-10-06

Baxter Healthcare Corporation is issuing a voluntary product recall for all unexpired lots of the 50mm 0.2 Micron Filter (Product Code: H93835) due to the potential for the filter membrane layer to be missing and for particulate matter to be present.

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →