FDA PRODUCT CODE · CLASS II

GWQ — Full-Montage Standard Electroencephalograph

Acquire, display, store, and archive electroencephalographic signals from the brain using a full montage array (i.e., 16 or more electrodes) and user-specified locations

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 882.1400
MEDICAL SPECIALTY
Neurology
REVIEW PANEL
NE
— Implant— Life-sustaining— GMP-exempt✓ Third-party review eligible
2
510(K)/PMA CLEARANCES
510k: 2
13
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is not implantable

SPEQ READINGNo implant-specific expectations attach from the classification itself.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is a candidate for a third party review program

SPEQ READINGAn Accredited Person may conduct the 510(k) review instead of FDA. FDA still makes the clearance decision — third-party review changes who reviews, never who decides.

Other product codes under 21 CFR 882.1400

One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.

21 CFR 882.1400 — Electroencephalograph

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

(b) Classification. Class II (performance standards).

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 882.1400 on eCFR →

Recent clearances

K261604Neuroelectrics Barcelona S.L.U. · 510k2026-06-13
K260455Zeto, Inc. · 510k2026-03-13

Recent recalls

Nihon Kohden America IncOpen, Classified · 2021-01-14

Medical device non-conformance to electrical safety standard (IEC 60601-1).

Natus Neurology IncTerminated · 2015-10-13

When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals.

Stellate SystemsTerminated · 2012-01-11

The following implied or stated indications in Stellate labeling with Harmoniae software have not yet been cleared by the Food and Drug Administration (FDA) for sale in the United States of America.

Respironics, Inc.Terminated · 2008-04-24

Foreign material: Glass fragments may present in the plastic bag material used to ship components.

Nicolet Biomedical Div of Viasys HealthcareTerminated · 2008-01-03

Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.

Bio-Logic Systems CorpTerminated · 2007-11-03

Bio-logic System Corp Camera Pole may weaken and detach at the base of the telescoping camera pole during transport causing pole to break at the welded base and fall.

Bio-Logic Systems CorpTerminated · 2007-04-20

Battery and Battery Compartment; incorrect placement in the Netlink Traveler unit which caused a short in the wiring of the battery pack and overheating.

Nicolet Biomedical Div of Viasys HealthcareTerminated · 2007-02-15

A software anomaly exists in the 5.20.1038 NicoletOne LTM, Sleep, nEEG, vEEG and ICU Monitor systems using the M, and/or C series amplifiers. If this anomaly occurs, the system will display the data for channel one in all channels on NicoletOne LTM, Sleep, nEEG, vEEG or ICU Monitor systems.

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →