FDA PRODUCT CODE · CLASS II

MBH — Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 888.3565
MEDICAL SPECIALTY
Orthopedic
REVIEW PANEL
OR
✓ Implant— Life-sustaining— GMP-exempt— Third-party review eligible
21
510(K)/PMA CLEARANCES
510k: 21
25
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is implantable

SPEQ READINGImplant status raises the evidentiary bar on biocompatibility and on traceability from finished device back to material lot.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is not a candidate for a third party review program

SPEQ READINGFDA reviews any premarket notification for this code directly.

21 CFR 888.3565 — Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

(b) Classification. Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance. [68 FR 14137, Mar. 24, 2003]

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 888.3565 on eCFR →

Recent clearances

K261275Corin, Ltd. · 510k2026-09-17
K261462Smith & Nephew, Inc. · 510k2026-09-08
K253728Microport Orthopedics, Inc. · 510k2026-08-21
K251840Zimmer, Inc. · 510k2026-05-28
K253189Zimmer, Inc. · 510k2026-03-09
K253328Medacta International S.A. · 510k2026-02-26
K253637Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 510k2026-02-18
K253239Signature Orthopaedics Pty, Ltd. · 510k2025-12-23

Recent recalls

MicroPort Orthopedics Inc.Open, Classified · 2024-12-18

Femoral implant porous size 5 right primary and femoral implant non-porous size 6 left primary have correct markings on devices, but may have incorrect packaging.

Encore Medical, LPOpen, Classified · 2023-12-07

Due to incorrect product/device within packaging.

Encore Medical, LPOpen, Classified · 2023-11-07

Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.

Encore Medical, LPOpen, Classified · 2020-10-07

There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.

Zimmer Biomet, Inc.Terminated · 2016-02-22

LDPE bag containing the implant adheres to the highly polished implant surface.

Stryker Howmedica Osteonics Corp.Terminated · 2013-04-12

With regards to ShapeMatch Cutting Guides using MRI Imaging, Stryker Orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing ShapeMatch Cutting Guides.

DePuy Orthopaedics, Inc.Terminated · 2013-02-15

DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be

DePuy Orthopaedics, Inc.Terminated · 2013-02-15

DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be