FDA PRODUCT CODE · CLASS III

LMH — Implant, Dermal, For Aesthetic Use

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class III
MEDICAL SPECIALTY
Unknown
REVIEW PANEL
SU
✓ Implant— Life-sustaining— GMP-exempt— Third-party review eligible
16
510(K)/PMA CLEARANCES
pma: 16
2
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

PMA

SPEQ READINGPremarket approval under 21 CFR Part 814. FDA reviews safety and effectiveness directly rather than by comparison, so clinical evidence is normally expected and the quality system is inspected before approval rather than after marketing.

DEVICE CLASS

Class III (high risk): general controls and Premarket Approval (PMA)

SPEQ READINGGeneral controls are insufficient to assure safety and effectiveness, so approval rests on direct evidence rather than equivalence.

MALFUNCTION REPORTING

Device is not eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGIndividual malfunction reporting under 21 CFR Part 803 applies. The complaint-handling system has to produce a decision per event, not per period.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is implantable

SPEQ READINGImplant status raises the evidentiary bar on biocompatibility and on traceability from finished device back to material lot.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is not a candidate for a third party review program

SPEQ READINGFDA reviews any premarket notification for this code directly.

Recent clearances

P250021Croma-Pharma GmbH · pma2026-10-08
P140029Q-Med AB · pma2026-09-18
P030050Q-Med AB · pma2026-09-17
P250020Merz North America, Inc. · pma2026-09-10
P170002Teoxane S.A. · pma2026-08-31
P250016Merz North America, Inc. · pma2026-08-31
P050047Allergan · pma2026-08-28
P110033Allergan · pma2026-08-28

Recent recalls

Merz North America, Inc.Terminated · 2015-12-03

The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.

Merz Aesthetics, Inc.Terminated · 2011-11-01

There is a potential leakage of product that may occur with Merz Aesthetics 1.5cc syringes of Radiesse Dermal Filler. When using the syringe, Health Care Professionals should be alerted to the potential for product that my leak pass the plunger at the proximal end of the syring. The defect may res

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →