FDA PRODUCT CODE · CLASS II

ITI — Wheelchair, Powered

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. FDA interprets “mobility to persons restricted to a sitting position” to mean the device type is intended to provide mobility to individuals who have mobility impairments and require an assistive device for mobility. NOTE: This type of device is not intended to climb stairs (see product code: IMK).

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 890.3860
MEDICAL SPECIALTY
Physical Medicine
REVIEW PANEL
PM
— Implant— Life-sustaining— GMP-exempt✓ Third-party review eligible
81
510(K)/PMA CLEARANCES
510k: 81
25
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is not implantable

SPEQ READINGNo implant-specific expectations attach from the classification itself.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is a candidate for a third party review program

SPEQ READINGAn Accredited Person may conduct the 510(k) review instead of FDA. FDA still makes the clearance decision — third-party review changes who reviews, never who decides.

21 CFR 890.3860 — Powered wheelchair

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

(b) Classification. Class II (performance standards).

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 890.3860 on eCFR →

Recent clearances

K261200Anhui JBH Medical Apparatus Co., Ltd. · 510k2026-10-06
K261079Anhui JBH Medical Apparatus Co., Ltd. · 510k2026-10-06
K262241Xuzhou Heabens Medical Technology Co., Ltd. · 510k2026-09-29
K262582Suzhou Green Pine Medical Co., Ltd. · 510k2026-09-23
K262054Yongkang Joymax Medical Technology Co., Ltd. · 510k2026-09-16
K260229Yongkang Hengyi Vehicle Co., Ltd. · 510k2026-09-08
K260053Beijing OrionStar Technology Co., Ltd. · 510k2026-09-08
K261354Foshan Sowell Medical Technology Co., Ltd. · 510k2026-08-21

Recent recalls

Permobil, AbTerminated · 2017-10-03

There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.

Permobil, AbTerminated · 2017-10-03

There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.

Pride Mobility Products CorpTerminated · 2017-01-03

The interface between the TB3 back and tilt base can separate when tilted beyond the specified angle.

Invacare CorporationTerminated · 2009-12-29

For some Invacare chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may do one or more of the following: (1) veer to one side; (2) rock forward onto its front riggings; (3) drive in an unintended c

Invacare CorporationTerminated · 2009-01-27

Invacare has become aware that on some chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may veer to one side or it may rock forward onto its front riggings.

Invacare CorporationTerminated · 2007-07-04

Software error may permit the chair to recline too far.

Eagle Parts and ProductsTerminated · 2006-10-19

The wheel hub bolts may loosen resulting in the wheels coming off.

Eagle Parts and ProductsTerminated · 2006-10-19

The wheel hub bolts may loosen resulting in the wheels coming off.

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →