FDA PRODUCT CODE · CLASS II

EHD — Unit, X-Ray, Extraoral With Timer

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 872.1800
MEDICAL SPECIALTY
Dental
REVIEW PANEL
RA
— Implant— Life-sustaining— GMP-exempt✓ Third-party review eligible
9
510(K)/PMA CLEARANCES
510k: 9
4
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is not implantable

SPEQ READINGNo implant-specific expectations attach from the classification itself.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is a candidate for a third party review program

SPEQ READINGAn Accredited Person may conduct the 510(k) review instead of FDA. FDA still makes the clearance decision — third-party review changes who reviews, never who decides.

Other product codes under 21 CFR 872.1800

One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.

21 CFR 872.1800 — Extraoral source x-ray system

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

(b) Classification. Class II.

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 872.1800 on eCFR →

Recent clearances

K260795Changzhou Sifary Medical Technology Co., Ltd. · 510k2026-09-14
K260367PDX Co., Ltd. · 510k2026-05-27
K253864Dental Imaging Technologies Corporation · 510k2026-03-20
K252110Guangzhou Ajax Medical Equipment Co., Ltd. · 510k2026-03-12
K254018Guilin Refine Medical Instrument Co., Ltd. · 510k2026-02-19
K252909Ningbo Runyes Medical Instrument Co., Ltd. · 510k2026-02-03
K251438Guilin Woodpecker Medical Instrument Co., Ltd. · 510k2025-09-04
K242185Raypia Co., Ltd. · 510k2024-12-04

Recent recalls

Panoramic Rental Corp.Terminated · 2018-01-09

The lift motor used to raise and lower the overhead chassis will wear if not properly maintained according to the maintenance schedule.

Gendex CorpTerminated · 2014-10-02

Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm.

Philips Medical Systems Sales & Service Region No. AmericaTerminated · 2004-07-20

The cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.

Panoramic Rental Corp.Terminated

Panoramic Corporation is performing a voluntary Medical Device Correction on certain PC-1000 X-ray systems due to the potential for a safety nut to be ineffective, which could result in blunt force trauma that may impact the head, neck or shoulders.

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →