1,158 matching recalls
DateClassFirmReasonQuality domain
2026-05-06Class IIPhilips North AmericaZ-1960-2026The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
2026-05-06Class IIMedline Industries, LPZ-2016-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIMedline Industries, LPZ-2032-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIElekta, Inc.Z-2006-2026Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
2026-05-06Class IIMedline Industries, LPZ-2009-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIImmuno-Mycologics, IncZ-1984-2026Neutralization Buffer may contain contaminants
2026-05-06Class IIAbbott Point Of Care Inc.Z-2001-2026Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.Risk Management & CAPA
2026-05-06Class IIPhilips North AmericaZ-1955-2026The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
2026-05-06Class IIMedline Industries, LPZ-2042-2026Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.Environmental Monitoring
2026-05-06Class IIDEPUY (IRELAND)Z-1993-2026The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.Environmental Monitoring
2026-05-06Class IIPhilips North AmericaZ-1957-2026The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
2026-05-06Class IIStraumann USA LLCZ-1995-2026Includes an incorrect screw seat interface.Documentation & Change Control
2026-05-06Class IIMedline Industries, LPZ-2038-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIMedline Industries, LPZ-2043-2026Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.Environmental Monitoring
2026-05-06Class IIMedline Industries, LPZ-2027-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IISiemens Healthcare Diagnostics, Inc.Z-1980-2026Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
2026-05-06Class IBoston Scientific CorporationZ-1770-2026Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
2026-05-06Class IIBeta Bionics, Inc.Z-1996-2026insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.Data Integrity
2026-05-06Class IIPhilips North AmericaZ-1958-2026The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
2026-05-06Class IIMedline Industries, LPZ-2029-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIMedline Industries, LPZ-2011-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIMedline Industries, LPZ-2022-2026During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.Environmental Monitoring
2026-05-06Class IIDEPUY (IRELAND)Z-1991-2026The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.Environmental Monitoring
2026-05-06Class IIDEPUY (IRELAND)Z-1986-2026The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.Environmental Monitoring
2026-05-06Class IIMedline Industries, LPZ-2039-2026Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.Environmental Monitoring