← FDA ENFORCEMENT
Z-2743-2026Class II

Daavlin Distributing Company

Recalling firm
Daavlin Distributing Company
Product type
Device
Status
Ongoing
Report date
2026-07-22
Distribution
US Nationwide distribution.
Firm state
OH
Product quantity
26 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Documentation & Change Control
Reason for recall

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Product description

Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Code / lot info

Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.

Other recalls SPEQ maps to Documentation & Change Control
2026-07-22Class IIFayus Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.
See all Documentation & Change Control recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.