Z-2638-2026Class II

Baxter Healthcare Corporation

Recalling firm
Baxter Healthcare Corporation
Product type
Device
Status
Ongoing
Report date
2026-07-08
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota, Maryland, and Pennsylvania and the countries of Canada, Hong Kong, India, Japan, Korea, Philippines, Taiwan, Thailand, Vietnam, Austria, Belgium, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iraq, Israel, Italy, Jordan, Kenya, Lebanon, Libya, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovakia, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arabia Emrites, Ukraine, Argentina, Brazil, Chile, Colombia, Costa Rica, Mexico, and Peru.
Firm state
IL
Product quantity
13 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.

Product description

Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables

Code / lot info

Lot Code: Product Code: 1739994 UDI-DI: 00887761972780 Serial Numbers: 109931797, 109931798, 109931799, 109943588, 109943590, 109948550, 109964963, 109964964, 109964969, 109964970, 109964971, 109980700, and 110006694