← FDA ENFORCEMENT
Z-2581-2026Class II

Medline Industries, LP

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-07-01
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Firm state
IL
Product quantity
113,843 kits
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Documentation & Change Control
Reason for recall

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Product description

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAM PACK-LF DYNJ21574J JUDKINS PACK DYNJ51126

Code / lot info

DYNJ21574J UDI-DI 10889942153497 Lots 21ALA586 21CLA398 21ELA408 21FLA144 21FLA788 21HLA420 21ILA431 DYNJ51126 UDI-DI 10889942122820 Lots 21ADA917

Other recalls SPEQ maps to Documentation & Change Control
2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class IIFayus Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.