Hologic, Inc

Recalling firm
Hologic, Inc
Product type
Device
Status
Ongoing
Report date
2026-05-06
Distribution
Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
Firm state
MA
Product quantity
1200 units
Initiated by
Voluntary: Firm initiated
Reason for recall

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Product description

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Code / lot info

Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;