Recalling firm
Remel Europe Ltd.
Product type
Device
Status
Ongoing
Report date
2026-04-29
Distribution
Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
Firm state
N/A
Product quantity
117 US; 950 OUS
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental MonitoringReason for recall
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
Product description
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Code / lot info
UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858
Other recalls SPEQ maps to Environmental Monitoring