← FDA ENFORCEMENT
Z-1922-2026Class II

Beckman Coulter, Inc.

Recalling firm
Beckman Coulter, Inc.
Product type
Device
Status
Ongoing
Report date
2026-04-29
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bangladesh, Bolivia (Plurinational State of), Bosnia and Herzegovina, Brazil, Canada, China, C¿te d'Ivoire, Croatia, Czechia, Denmark, Egypt, El Salvador, Ethiopia, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Kosovo, Lebanon, Libya, Malawi, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, Spain, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates, Uzbekistan, Viet Nam, Yemen.
Firm state
MN
Product quantity
2068 US, 2830 OUS
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Documentation & Change Control
Reason for recall

Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.

Product description

Access Total T4 Calibrator, Catalog No. 33805

Code / lot info

UDI-DI 15099590225872 Lots 538729, 538815, 538898, 539107, 539243

Other recalls SPEQ maps to Documentation & Change Control
2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class IIFayus Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.
See all Documentation & Change Control recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.