Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Catalog Number: 82-0004; UDI-DI: 00850355006031; ALL lots and ALL sublots;
Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.