All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
MiniMed 620G Insulin Pump (MMT-1750)
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 620G Insulin Pump: 643169559745, 643169721319, 763000375461, 763000253189: Instr, Startup Guide, 620G 4.10, EN(6025951-022 1); INSTR., NGP 620G, BOLUS WIZARD, EN(6026006-023 1); Instr., 620G, Sftwr. 4.10, EN.(6026006-024 1); INSTRUCTIONS MMT-746EN PUMP 1510/1710 EN(7005933-023 1); DOC PKG MMT-746EN NGP 1710 EN(7005933-024 1); Startup Guide Assembly MMT-755EN(620G(M982438A021 1).