← FDA ENFORCEMENT
Z-1639-2026Class II

Olympus Corporation of the Americas

Recalling firm
Olympus Corporation of the Americas
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Firm state
PA
Product quantity
164 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Product description

Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.

Code / lot info

Model: 3006; UDI: 00821925035898; Lot Numbers: All unexpired;

Other recalls SPEQ maps to Risk Management & CAPA
2026-07-22Class IAbiomed, Inc.2026-07-22Class IAbiomed, Inc.2026-07-22Class IAbiomed, Inc.2026-07-22Class IAbiomed, Inc.
See all Risk Management & CAPA recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.