← FDA ENFORCEMENT
Z-1981-2026Class I

American Contract Systems Inc

Recalling firm
American Contract Systems Inc
Product type
Device
Status
Ongoing
Report date
2026-05-13
Distribution
US distribution to MN. No OUS distribution.
Firm state
OH
Product quantity
7,311 kits
Initiated by
Voluntary: Firm initiated
Reason for recall

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Product description

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

Code / lot info

ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911