Medline Industries, LP
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
1. UDI-DI 10195327501259 (EA) 40195327501250 (CS) LOTS 24DBW065 24EBV301 24HBB492 24JBW143 25ABG413 25CBH359 25DBD213 25GBP569 25HBQ088 25KBF725 26ABE643 2. UDI-DI 10195327371029 (EA) 40195327371020 (CS) LOTS 24FBM018 24GBL319
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →