← FDA ENFORCEMENT
Z-1708-2026Class II

Medline Industries, LP

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-04-08
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Firm state
IL
Product quantity
14,379 kits total
Initiated by
Voluntary: Firm initiated
Reason for recall

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Product description

Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915

Code / lot info

1. UDI-DI 10195327501259 (EA) 40195327501250 (CS) LOTS 24DBW065 24EBV301 24HBB492 24JBW143 25ABG413 25CBH359 25DBD213 25GBP569 25HBQ088 25KBF725 26ABE643 2. UDI-DI 10195327371029 (EA) 40195327371020 (CS) LOTS 24FBM018 24GBL319