← FDA ENFORCEMENT
Z-1640-2026Class II

Olympus Corporation of the Americas

Recalling firm
Olympus Corporation of the Americas
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Firm state
PA
Product quantity
444 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Product description

Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation during an endoscopic or open surgical procedure.

Code / lot info

Model: PK-CF0533; UDI: 00821925035867; Lot Numbers: All unexpired;