Recalling firm
Orthalign, Inc
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
US Nationwide distribution in the state of UT.
Firm state
CA
Product quantity
7 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & QualificationReason for recall
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Product description
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System
Code / lot info
Lot # 25092204BB/ UDI:00810832032434
Other recalls SPEQ maps to Validation & Qualification
See all Validation & Qualification recalls →Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.