← FDA ENFORCEMENT
Z-2719-2026Class II

Olympus Corporation of the Americas

Recalling firm
Olympus Corporation of the Americas
Product type
Device
Status
Ongoing
Report date
2026-07-22
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.
Firm state
PA
Product quantity
1,156 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.

Product description

Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.

Code / lot info

Model Number: 188A. UDI-DI: 00821925001077. All Lot Numbers

Other recalls SPEQ maps to Environmental Monitoring
2026-07-22Class IFrutas y Hortalizas del Sur S.A.2026-07-22Class IIGELLERT GLOBAL GROUP2026-07-22Class IICareFusion 213, LLC2026-07-22Class IIChiesi USA, Inc.
See all Environmental Monitoring recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.