← FDA ENFORCEMENT
Z-2703-2026Class II

Siemens Healthcare Diagnostics, Inc.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2026-07-15
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bangladesh, Bolivia, Brazil, Bulgaria, Burkina¿Faso, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Democratic Republic Congo, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Poland, Portugal, Qatar, Republic of Korea, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Thailand, Turkey, Turkmenistan, U.A.E., Uganda, United¿Kingdom, Uruguay, Vatican City, Vietnam.
Firm state
NY
Product quantity
24,077 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.

Product description

Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.

Code / lot info

Material Number: 11097536. UDI-DI: 00630414220505. All Lot Numbers.

Other recalls SPEQ maps to Validation & Qualification