A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set Code 1000441, containing Kit, 23Fr Introducer, 11cm, Sterile, Introducer Product Code 2000342. heart pump and accessories
Pump Set Code 1000441; Introducer Product Code 2000342. GTIN: 00813502012910. All current batches in field.