Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System
REF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095
Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.