← FDA ENFORCEMENT
Z-2553-2026Class II

GE Medical Systems SCS

Recalling firm
GE Medical Systems SCS
Product type
Device
Status
Ongoing
Report date
2026-07-01
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
Firm state
N/A
Product quantity
340 servers
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Documentation & Change Control
Reason for recall

GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

Product description

GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

Code / lot info

REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384