Recalling firm
Smith & Nephew, Inc.
Product type
Device
Status
Ongoing
Report date
2026-06-24
Distribution
International distribution to the countries of BE, DE, IT.
Firm state
TN
Product quantity
5
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPAReason for recall
Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.
Product description
TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
Code / lot info
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
Other recalls SPEQ maps to Risk Management & CAPA
See all Risk Management & CAPA recalls →Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.