Recalling firm
GE Medical Systems, LLC
Product type
Device
Status
Ongoing
Report date
2026-06-17
Distribution
Worldwide - US Nationwide distribution.
Firm state
WI
Product quantity
5 units
Initiated by
Voluntary: Firm initiated
Reason for recall
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Product description
Allia IGS Pulse angiographic X-ray system
Code / lot info
UDI-DI: 00195278827647; Serial Numbers: DVMSS2500001HL, DVMSS2500002HL, DVMSS2500003HL, DVMSS2500004HL, DVMSS2500005HL, DVMSS2500006HL, DVMSS2600001HL, DVMSS2600002HL