Z-2284-2026Class II

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product type
Device
Status
Ongoing
Report date
2026-06-10
Distribution
US Nationwide distribution.
Product quantity
16 units
Reason for recall

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Product description

Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Code / lot info

UDI (01)00884838099203(21)3, (01)00884838099203(21)7, (01)00884838085275(21)131, (01)00884838099203(21)54, (01)00884838085275(21)84, (01)00884838099203(21)6, (01)00884838099203(21)93, (01)00884838085275(21)132, (01)00884838099203(21)69, (01)00884838099203(21)9, (01)00884838099203(21)115, (01)00884838099203(21)85, (01)00884838099203(21)19, (01)00884838099203(21)49, (01)00884838099203(21)108, (01)00884838085275(21)108.

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.