Recalling firm
Stryker Corporation
Product type
Device
Status
Ongoing
Report date
2026-06-10
Distribution
US Nationwide distribution in the state of Indiana.
Firm state
MI
Product quantity
1 unit
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPAReason for recall
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
Product description
Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System
Code / lot info
GTIN 07613327634914, Serial Number 2521008873
Other recalls SPEQ maps to Risk Management & CAPA