← FDA ENFORCEMENT
Z-2212-2026Class I

KAYSERBETTEN GMBH & CO. KG

Recalling firm
KAYSERBETTEN GMBH & CO. KG
Product type
Device
Status
Ongoing
Report date
2026-06-10
Distribution
Distribution US Nationwide and Canada.
Product quantity
372 units
Initiated by
Voluntary: Firm initiated
Reason for recall

If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).

Product description

Pediatric care bed; Product Designation: KayserBett IDA;

Code / lot info

Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6351, 6355, 6795, 6796, 7176, 7177, 7178, 7179, 7180, 7181, 7182, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7617, 7618, 7619, 7620, 7621, 7622, 7623, 8242, 8243, 8244, 8245, 8588, 8589, 8590, 8591, 8592, 8593, 8594, 8595, 8596, 8597, 8598, 8599, 8600, 8601, 8602, 8603, 8608, 8609, 8691, 8704, 8912, 8921, 8922, 8923, 8924, 8925, 8926, 8927, 8928, 8929, 8930, 9725, 9726, 9727, 10350, 10351, 10352, 10819, 10820, 10821, 10822, 10823, 10824, 10825, 11186, 11367, 11368, 11369, 11370, 11371, 11372, 11797, 11798, 11799, 11800, 11801, 11802, 11803, 11804, 11805, 11806, 11807, 11808, 12247, 12248, 12249, 12250, 12251, 12252, 12253, 12254, 12255, 12256, 12257, 12397, 12398, 12399, 12782, 12783, 12963, 12964, 12965, 12966, 12967, 12968, 12969, 12970, 12971, 12972, 13486, 13487, 13488, 13489, 13490, 13491, 13492, 13493, 13494, 13495, 13496, 13497, 13912, 13913, 13914, 13915, 13916, 13917, 13918, 14942, 14944, 14952, 14953, 15912, 15913, 15914, 16403, 16404, 16405, 16406, 16407, 16408, 16409, 16410, 16411, 16984, 16985, 16986, 16987, 16988, 16989, 16990, 16991, 16992, 16993, 16994, 17577, 17578, 17579, 17580, 17581, 17582, 17583, 17584, 17585, 17586, 17587, 18494, 19112, 19113, 19114, 20151, 20152, 20153, 20154, 20155, 20156, 20157, 20158, 20963, 20964, 20965, 20966, 20967, 20968, 20969, 20970, 20972, 21192, 21193, 21194;

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.