← FDA ENFORCEMENT
Z-2203-2026Class II

Becton, Dickinson and Company

Recalling firm
Becton, Dickinson and Company
Product type
Device
Status
Ongoing
Report date
2026-05-27
Distribution
Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.
Firm state
CA
Product quantity
183
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

Product description

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P

Code / lot info

Model-UDI-DI(Expiration): SGPT755P/00690103219388/65869018(6/24/2026), 66010501(9/23/2026), 67122784(10/28/2027); SGPT54/00690103219289/65956343(8/19/2026); SGPT64P/00690103219371

Other recalls SPEQ maps to Risk Management & CAPA
2026-07-22Class IAbiomed, Inc.2026-07-22Class IAbiomed, Inc.2026-07-22Class IAbiomed, Inc.2026-07-22Class IAbiomed, Inc.
See all Risk Management & CAPA recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.