Erbe Medical, LLC

Recalling firm
Erbe Medical, LLC
Product type
Device
Status
Ongoing
Report date
2026-05-13
Distribution
US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
Firm state
AZ
Product quantity
260,962
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

Product description

ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208

Code / lot info

REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249; 20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243; 20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367; 20325-207/04065655000689/W4469939, W4469940, W4469941; 20325-248/04065655001006/W4467050, W4467051, W4467055, W4467056, W4467682, WO470898, WO470899, WO470900, WO470901, WO470902, WO470903, WO471252, WO471253, WO471254, WO471255, WO471256; 20325-208/04065655001099/W4467040, W4468271, WO468760