Z-1935-2026Class I

Cook Incorporated

Recalling firm
Cook Incorporated
Product type
Device
Status
Ongoing
Report date
2026-05-13
Distribution
US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay
Firm state
IN
Product quantity
271
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

Product description

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

Code / lot info

UDI-DI 00827002473089 Lots 16858459 16869145 16857138 16763324 16750208 16752733 16821053 16713191