← FDA ENFORCEMENT
Z-1880-2026Class II

Medline Industries, LP

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-04-29
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Firm state
IL
Product quantity
270,311 total
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Documentation & Change Control
Reason for recall

Unapproved design changes to the products outside of the 510(k) clearance.

Product description

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262

Code / lot info

DYNJ0101278F UDI-DI 10193489352351 (EA) 40193489352352 (CS) LOTS 22CDB133 22EDA975 22FDA831 22HDC403 DYNJ0101278G UDI-DI 10195327286477 (EA) 40195327286478 (CS) LOTS 22KDB109 22LDA476 23CDA805 23DDB546 23EDC278 23FDC023 23JDA625 23KDB893 24ADB642 24BDB842 24CDB908 24GDA043 24IDA026 DYNJ0101278I UDI-DI 10198459119477 (EA) 40198459119478 (CS) LOT 24JDC169 DYNJ56262 UDI-DI 10889942624331 (EA) 40889942624332 (CS) LOT 24ABO125 ***Updated 5/20/26 - The following lots were included in the customer letter but not in the FDA submission.*** DYNJ0101278F lots 21ADC024 21CDA355 21EDA527 21GDC529 21IDA040 21IDA786 21JDB214 21KDC443 22ADC117 22GDA789; DYNJ0101278G lots 23HDC100; DYNJ0101278I lots 25AME306 25AMI354 25BMG147 25BMG147A

Other recalls SPEQ maps to Documentation & Change Control
2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class IIFayus Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.2026-07-22Class ID'Dioses Fruit Pops, Inc.
See all Documentation & Change Control recalls →

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.