Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Philips Spectral CT on Rails. Model Number: 728334.
Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.