Z-1641-2026Class II

Olympus Corporation of the Americas

Recalling firm
Olympus Corporation of the Americas
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Firm state
PA
Product quantity
2648 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Product description

Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.

Code / lot info

Model: HACF0533; UDI: 00821925036390; Lot Numbers: All unexpired;