← FDA ENFORCEMENT
Z-1598-2026Class II

I.T.S. GmbH

Recalling firm
I.T.S. GmbH
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Product quantity
510 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Product description

I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.

Code / lot info

1. Article Number (Lot Numbers): 21604-6 (643/052301, 643/0652, 643/092201), 21604-9 (643/032311, 643/062308, 643/062309, 643/112310, 643/122201). 2. Article Number (Lot Numbers): 21605-11 (643/072314, 643/102207), 21605-6 (643/0654, 643/072306, 643/072316, 643/072338, 643/102208, 643/102212, 643/102217, 643/112306, 643/122202), 21605-9 (643/0643, 643/0662, 643/072307, 643/072339, 643/112309, 643/122203). 3. Article Number (Lot Numbers): 21105-6 (86/876787), 21105-8 (86/876788). 4. Article Number (Lot Numbers): 21104-4 (679/052234, 679/082305, 86/036334), 21104-5 (679/0722101, 679/082306, 86/036335), 21104-6 (679/012307, 86/876760), 21104-7 (679/082307, 86/036336, 86/876781), 21104-8 (679/082308, 86/876783), 21104-9 (86/876904). 5. Article Number (Lot Numbers): 21101-5 (86/876637), 21101-6 (679/121605, 00/1105-12, 26/F41923, 86/876638, 86/903610), 21101-7 (770/201905, 00/0518-06, 679/042303, 86/876640), 21101-8 (86/876721, 679/062224), 21101-9 (679/122264, 86/876727), 21101-11 (86/876633), 21101-13 (86/876635). 6. Article Number (Lot Numbers): 21102-6 (770/201905). 7. Article Number (Lot Numbers): 21103-8 (679/092216), 21103-9 (679/062215, 86/876758), 21103-10 (679/092219, 86/876741), 21103-11 (679/062214), 21103-12 (86/876743), 21103-13 (679/022303, 86/876745).

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.