← FDA ENFORCEMENT
Z-1588-2026Class I

Navajo Manufacturing Company

Recalling firm
Navajo Manufacturing Company
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
Firm state
CO
Product quantity
28,457
Initiated by
Voluntary: Firm initiated
Reason for recall

If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.

Product description

Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd

Code / lot info

UPC: 0-24291-25607-6, UDI-DI: 00024291256076

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.