FDA PRODUCT CODE · CLASS II

ONG — Powered Laser Surgical Instrument With MicrobeamFractional Output

A LASER (Light Amplification by Stimulated Emission of Radiation) based device having coherence, collimated and typically monochromatic radiation. The laser output is in form of tiny beams in micro range. Typically, indicated to be used in medical specialties of general and plastic surgery and of dermatology/aesthetic to destroy, remove or coagulate tissue, generally soft tissue. Indication for Use can be general and/or specific. The classification regulation 21 CFR 878.4810 describes a device that is carbon dioxide or argon laser intended to cut, destroy, remove or coagulate tissue by the light.

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 878.4810
MEDICAL SPECIALTY
General, Plastic Surgery
REVIEW PANEL
SU
— Implant— Life-sustaining— GMP-exempt✓ Third-party review eligible
4
510(K)/PMA CLEARANCES
510k: 4
3
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is not implantable

SPEQ READINGNo implant-specific expectations attach from the classification itself.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is a candidate for a third party review program

SPEQ READINGAn Accredited Person may conduct the 510(k) review instead of FDA. FDA still makes the clearance decision — third-party review changes who reviews, never who decides.

Other product codes under 21 CFR 878.4810

One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.

21 CFR 878.4810 — Laser surgical instrument for use in general and plastic surgery and in dermatology

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification.

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.

(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

(b) Classification.

(1) Class II.

(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9. [53 FR 23872, June 24, 1988, as amended at 61 FR 1123, Jan. 16, 1996; 66 FR 38803, July 25, 2001]

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 878.4810 on eCFR →

Recent clearances

K252155AVAVA, Inc. · 510k2025-10-20
K244060Lutronic Corporation · 510k2025-07-31
K251031Shanghai Apolo Medical Technology Co., Ltd. · 510k2025-07-02
K250402AVAVA, Inc. · 510k2025-04-14

Recent recalls

IDS LTDOpen, Classified · 2025-01-07

The Smaxel Fractional CO2 Laser does not include a Certification label, and adequate instructions for safe operation of the device.

WINGDERM ELECTRO-OPTICS LTD.Open, Classified · 2024-12-05

Model WFB-01 sold to US customers before Aug 23, 2022 has been found to have a software bug (6 units). All WFB-01 models sold to US customers found to have labeling non-compliances (22 units)

Solta Medical, Inc.Terminated · 2011-02-01

The product has the potential to cause inadvertent laser firing which may lead to patient/operator injury if the problem recurs.

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →