Fresenius Medical Care Holdings, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
Model Number: RFP-400-G; UDI-DI: 00840861102839; 2. Model Number: RFP-401-G; UDI-DI: 00840861102853; 3. Model Number: RFP-402-G; UDI-DI: 00840861102822; 4. Model Number: RFP-404-G; UDI-DI: 00840861102914; 5. Model Number: RFP-407-G; UDI-DI: 00840861102846: 6. Model Number: RFP-456-G; UDI-DI: 00840861102945; All lots produced from January 13, 2025, through lot 26AG02019 on January 13, 2026
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →